DSI-7 BCI 2000 Package
Intuitive BCI package with Dry EEG
- Product No: NS-1071
- Manufacturer: Wearable Sensing
- EEG ERP BCI FFT
- Request Quote
Description
The DSI-7 is a complete, research-grade wireless EEG system designed for rapid application of 7 sensors at positions corresponding to locations within the international 10-10 system. Sensor locations can be customized on demand by the manufacturer. The system comprises ultra-high impedance active Dry Sensor Interface (DSI) sensors that function through hair, requiring no skin preparation or conductive gels. The sensors can be individually adjusted to optimize contact impedance. They are spring loaded to provide constant, comfortable contact pressure that mitigates movement artifacts seen during ambulation and are actively and passively shielded to prevent contamination from electrical artifacts.
The DSI-7's practicality, portability, and high signal quality opens the door to taking numerous applications that require rapid EEG recording. Such applications include:
- Augmented cognition
- Biometrics
- Brain-computer interfaces
- Cognitive stress or workload monitoring
- Neuroeconomics / Neuromarketing
- Neuroergonomics
- Neurofeedback
- Psychological Research
- Sports Peak-Performance training
System Features
- Adjustable to fit a wide range of headsizes
- Bluetooth wireless transmission with sampling at 300 Hz
- Comfortable for wearing continuously over 8 hours
- Continuous impedance monitoring
- Linked-Ears earclip mounted active sensors
- Portable design suitable for light ambulation in office or lab environment
- Rechargeable Li-ion batteries allow continuous operation
- Self-donned and ready to record EEG in under 1 minutes
- 7 sensors positioned at choice of F3, F4, C3, C4, Pz, P3, P4 locations of the 10-20 International System
- 4-bit digital trigger input
Technical Specifications
Sensor Locations | Seven (7) sensors at fixed locations of theInternational 10/20 System locations:F3, F4, C3, C4, Pz, P3, P4 + Linked Ears |
Common-Mode-Follower | One (1) Common Mode Follower Sensor at Pz is default reference (can be changed to LE in software) |
Ground | One (1) ground electrode at the last remaining location |
Headset | Easy to don, comfortable, and long-term wearable headset assembly with spring-loaded arms and pods for sensor support. Good for easy rapid use in office or laboratory environment. |
Amplifier/Digitizer | One (1) 7-channel amplifier and digitizer |
Wireless | Bluetooth |
Run-time | >12 hours |
Headsize fit | 52-62cm circumference |
Headset weight | ~350g |
Wireless range: | 10m |
Digital Inputs: | 4 bit input |
Analog/Digital Converter: | 16 bits |
Analog/Digital resolution: | 0.317 uV referred to input |
Sample Rate: | 300 Hz |
Bandwidth: | 0.003-150Hz |
Gain: | 60 |
Common mode rejection ratio: | > 120dB |
Channel Cross-talk: | < -70dB with sensors |
Input Impedance (1Hz): | 47Gohm |
Input bias current: | <25pA |
DC offset tolerance: | +/-200 mV |
Maximum input range: | 10mV pk-pk |
Noise (1-50Hz): | <3 uV pk-pk |
Please note: Important! This product is for research applications only. Not a medical device as defined in EU directive 93/42/EEC. Not designed or intended to be used for diagnosis or treatment of disease.